Instructions for Use V 1.9 RevD - 05/2013
6. Cleaning and Maintenance
• Pusha(new)cleaningcartridgeasfarasit
will go into the shaft (Fig. 58).
• Unscrewtheredcap(85) one turn and slide
the cartridge fully home.
• Pushthe securitylatch(86, Fig. 57) back in
front of the cartridge.
+
Instructions for storage (continued)
Fig. 59
72
69
Fig. 58
• Pushthe smalllever(69, Fig. 59) upwards
under the print head.
64
85
• Openthe hoodofthe printer
(4 in Fig. 31), then press the
CLEAN and LOADED keys simul-
taneously.
• Theprinthead (72) moves up-
ward to a position approx. 1 cm
(
1
/
2
inch) away from the sealing
lip (Fig. 59).
• The"InkEmpty"LED will go out and "88" will
appear on the display.
• PressCLEAN to rinse the print head; "00" will
appear on the display. The display will go out
when the cleaning process is complete.
The cleaning process with solvent
takes approx. 3.5 minutes.
ð