Leica-biosystems IPS Manuale Utente Pagina 61

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 60
59
Leica IP S
7. Troubleshooting
7.1 Malfunctions
+
Confirm elimination
of error
+
If the printer malfunctions during printing, a corresponding error
code is displayed and simultaneously the LED in the ERROR button
starts flashing.
How to eliminate the problem:
• Determinethecauseoftheerror—seeerrorlist(Chapter 7.3).
• Removeobstruction(s);openlidifrequired.
• Important!
Remove all slides which are still in the chute, in or next to the specimen
slide carrier or in the drying module!
These specimen slides should not be used again.
• ReclosethehoodandpresstheERROR button to confirm to the printer
that the source of the error has been eliminated.
• Theprinterthenverifieswhetherallprocessingpathsareunobstructed
and whether the source of the error has been eliminated.
• Iftherearestillsomeobstructionsleft,orifthesourceoferrorhasnot
been thoroughly eliminated, the printer displays another error message.
• Interruptedprintjobsareresumedwherepreviouslyleftoff.
• Ifanerrormessageisdisplayedseveraltimesalthoughallpossible
causes have been eliminated, a Reset should be carried out.
Reset:
• Pressandrelease LOADED and ERROR simultaneously.
• AResetrestorestheprintertothesamestateasdirectlyafterswitching
on. All print jobs in the print queue are deleted.
• IfthesameerrorcontinuestobedisplayedevenafteraReset,switchthe
printer off via the power switch (back panel) and, after a short waiting
period of approx. 30 sec., switch it back on. If this does not eliminate the
problem either, call Leica Technical Service.